Seite 1 - Discontinued Instrument

CEL-281COMPUTING NOISEDOSIMETEROperator's HandbookDocument 060112Issue: 4BMarch 1996archive DocumentforDiscontinued Instrument

Seite 2

920066920067If the display blinks,remains blank, or LOWBAT is indicated: eitherthere is insufficientpower or the batteriesare inserted incorrectly(see

Seite 3 - Contents Overview

Provided the CEL-281 is not in calibration delay mode or calibrationmode, and the display is switched ON, it can be used as follows to ob-tain and dis

Seite 4

4. Change the time weighting in use as detailed inSection 4.7.1.5. Provided the instrument is not logging data, press the keyto change the frequency w

Seite 5

criterion levels can be accepted for use, i.e. criterion levels≥threshold level as described in Section 2.2.2 and shown inTable 1.9. Press the key aga

Seite 6

enter the fully operational state with the display switched ON,(with or without data logging) as described in Chapter 4.If the instrument has just bee

Seite 7

If overloading has occurred, the display will indicate the per-centage of the elapsed time spent overloaded.6. Press the key to obtain underrange info

Seite 8

9. Press the key to see the earliest time or date at which the200 Pa level was exceeded.10. Press the key to display the time.11. Press the key to dis

Seite 9 - 1.1 Check Batteries

2. Press the key to enter the distribution operational area.If data is being logged, logging will continue uninterrupted, butthe display will change t

Seite 10 - 1.2 Load Batteries

4. Press the key to display the relevant cumulative percentvalues.5. Press the key to display the relevant probability percentvalues.6. Then press eit

Seite 11 - 1.3.1 Quick Start

1. Load and/or check batteries, perform a field accuracy check(acoustic calibration), select time and frequency weightings, andenter the fully operati

Seite 12 - 1.3.2 Acoustic Calibration

1 min. slow max Maximum SPL taken over a 1 minutesampling period with slow response,1 min. imp. max Maximum SPL taken over a 1 minutesampling period w

Seite 13 - Getting Started

4. Load four AAA orLR03 batteriesinto the holdersupplied, accord-ing to polaritiesshown on thebottom of theholder, then in-sert it into thebattery com

Seite 14 - 1.3.4 Set Frequency Weighting

5. Press the key to display the relevant dB level.6. Press the or keys to select respectively, the next lateror earlier cell in the currently displaye

Seite 15 - 1.5 Read L

The dosimeter is now operating as a sound level meter, indi-cating the instantaneous A-weighted Fast SPL on the main dis-play, with the current time c

Seite 16 - 1.6 Read Pa

The main display is now indicating A-weighted Slow SPL.13. Press the key a sixth time.The main display is now indicating A-weighted Fast Max, and ifth

Seite 17 - 1.7 Read %Dose & P%Dose

5.6.3 Measuring %Dose and P%Dose1. Switch the CEL-281 ON.2. Slide the CEL-6012 Keypad on to the dosimeter.3. Wait about 30 seconds until TC= , , or is

Seite 18 - Getting Started_

(Q=3 is normally used in ISO measurements of noise dose,while Q=5 is specified for USA OSHA estimates of the risk ofhearing loss.)8. Inspect the curre

Seite 19

13. Press the key again to see the result of dose calculationusing the new threshold level.5.6.4 Finding Amount of Time Spent Above aSpecified Level1.

Seite 20

6. Press the or keys to step forwards or backwardsthrough the series of cells to reach the cell corresponding to thespecified level.7. Press the key t

Seite 21 - 2.1 Introduction

6. Hold down the key and press the key once.Depending on whether time or date was inspected last, themain display indicates the first time or date at

Seite 22 - Introduction_

6 PRINTING HARD COPY REPORTSIt is possible to download a "subset" of the contents of the CEL-281Dosimeter store to a dot matrix printer via

Seite 23 - _Introduction

3. Carefully remove the front cover. All circuitry is retained in therear cover.4. Set the two DIP switches to produce the required report format,refe

Seite 24 - 2.2 Explanation of Terms Used

4. If an incorrect time weighting for the required measurements isindicated at "TC=?", go to Section 1.3.3: Set Time Weighting.5. If an inco

Seite 25

Paper end: Valid,Auto linefeed: Off (The dosimeter sends a linefeed),Skip overperforation: Invalid,SLCT IN Internally fixed,Printer mode Centronics.6.

Seite 26

4. Either: If the dosimeter is switched ON and logging data, pressthe + key combination to stop logging, then go tostep 5.Or: If the dosimeter is curr

Seite 27

CEL 281 ISO ReportInstrument I.D. Number : 000000Subject : MachinistLocation : Machine Shop 2dd/mm/yyhh:mmLogging Calibration:01/01/9100:31Logging Sta

Seite 28

1401351301251201151101051009590858075708.45time------>1 min Leq920069dB Level -- -- -- >Figure 22: Portion of graphical plot from ISO hard copyP

Seite 29

7 TECHNICAL INFORMATION7.1 Schedule of PartsA complete "CEL-281 Computing Noise Dosimeter" consists only of:CEL-281 Noise Dosimeter, include

Seite 30

7.2 Replacing The MicrophoneCautionThis procedure should be performed only bysuitably qualified personnel.If the microphone or connecting cable become

Seite 31

The connector is keyed so that the plug can be inserted only thecorrect way.7. Carefully replace the upper circuit board.Locate the three 3-pin connec

Seite 32

10. Press the ON/OFF switch in to latch the instrument ON (seeSection 4.1).The instrument will go through its self checking program.11. Push the micro

Seite 33

pressure, the CEL calibrators generates a nominal 1 kHz signalwith a level of 114.0 dB at the microphone. The dosimeter dis-play should indicate 114.0

Seite 34

18. Using a slim screwdriver, reach through the hole in the uppercircuit board and adjust the trim potentiometer on the lowerboard (shown in Figure 23

Seite 35 - 3.1 Principle of Operation

5. Install the keypad.If the 20 s delay has elapsed before the cover and keypad havebeen installed, CAL stops blinking and the instrument becomesfully

Seite 36 - How It Works

7.3 SpecificationType:Noise Dosimeter according to ANSI S1.25-1978 Class 2A & 2C*,Personal Sound Exposure Meter according to IEC 1252:1993 and BS

Seite 37

Reference frequency:1kHz.Reference direction:Perpendicular to microphone diaphragm.Reference SPL & calibration integration time forsound exposure:

Seite 38 - 3.2 Technical Description

CEL-6225 Dust Cover has no measurable effect on acousticperformance of microphone.Electrical substitution of microphone input forcalibration:Not recom

Seite 39 - Digital Section

Stored & displayed parameters:Level information:Sound pressure level (dB) with Fast time weighting,Sound pressure level (dB) with Slow time weight

Seite 40

6 different criterion levels, and 5 different threshold levels.Memory capacity: 14 hours*.Time information:Date and time,Elapsed time,Proportion of ov

Seite 41

Digital output:CEL low-power Interface, 4800 baud.Dimensions:Overall dosimeter plus keypad:150 x 83 x 41 mm, 310 g complete with batteries and battery

Seite 42

8 MANUFACTURERS SERVICING &WARRANTY ARRANGEMENTSIn order to ensure its rigid conformity with the requirements of the speci-fication, this instrume

Seite 43

In addition, CEL Instruments has a NAMAS accredited laboratory No.0237 which is able to perform certain acoustic calibrations with traceabil-ity to U.

Seite 44

C-RETINFReason for Return/Nature of Fault DateCustomer ReferenceCustomer NameAddressPhone No.Contact NameInstrument TypeSerial NumberInstrument TypeSe

Seite 45

Service & Warranty_

Seite 46

WarningTo prevent damage, DO NOT EXERT FORCE onthe flexible crosspiece when removing themicrophone from the calibrator coupler.PULL ONLY on the rigid

Seite 47

≅≡≡IndexReference PageReference PageIndex

Seite 48

Reference PageReference PageIndex_

Seite 49

Reference PageReference PageIndex

Seite 50

Reference PageReference PageIndex_

Seite 51

Reference PageReference PageIndex

Seite 52

Reference PageReference PageIndex_

Seite 53

Reference PageReference PageIndex

Seite 54

Index_

Seite 55

1. Press the key to start data logging.The instrument starts to store data samples at the beginning ofthe next whole clock minute.2. Press the key to

Seite 56

4. Press the key once.5. If some other value than 3 is shown at "Q=?", press thekey enough times to obtain "Q=3".The main display

Seite 57

The main display now indicates sound exposure level in Pa h,and if the instrument contains stored data or is logging data, alevel will be given.1.7 Re

Seite 58

6. If necessary, view the current threshold level (at "Thr") by press-ing the key once.7. If necessary, press the key slowly sufficient time

Seite 59

6. Press the key once.The main display indicates the last time or date when the200 Pa level was exceeded.7. If time is indicated, press the key once t

Seite 60

Supplementary Information - July 1996Revised Microphone ArrangementsSince this handbook was prepared, a redesigned microphone arrange-ment has come in

Seite 61

Getting Started_

Seite 62

2 INTRODUCTION TO THE CEL-2812.1 IntroductionThe CEL-281 Computing Noise Dosimeter stores all of its parameters si-multaneously, so the operator may n

Seite 63

MealTime (hours)LevelLevelTime (hours)LevelTime (hours)LevelMachine noise atSLM like CEL-231 orvariations in level is bestMachine noise with largemeas

Seite 64

Health Authority regulations OSHA 1910.95 "Occupational Noise Expo-sure".The CEL-281 is intended to be worn by an individual so that its mic

Seite 65 - 3.4 Memory Functions

The keypad can be removed to eliminate the danger of accidentally re-setting the instrument or losing measurements. This will also preventunauthorised

Seite 66 - 3.5 Interface

Both terms are in common use, but the most recent international stand-ards specify sound exposure.The value of sound exposure measured (E) is normally

Seite 67

At levels below the First Action Level (1.0 Pa h - equivalent to 85 dB(A)for 8 hours) "reasonably practicable" noise reductions should be ma

Seite 68 - 3.6 Power Supply

Table 1: Valid pairs of criterion and threshold levelsℵℵℵℵℵℵℵℵℵℵℵℵℵℵℵℵℵℵ2.2.3 ThresholdMost sound exposure legislation requires that all noise measure

Seite 69

Results obtained by various researchers indicate that when the risk ofhearing damage is being considered, less emphasis need be placed onthe energy co

Seite 70 - 4 PREPARATION FOR USE

2.2.5 Noise DoseNoise dose is not an absolute measurement but a percentage of the rec-ommended maximum daily personal noise exposure.In this way, 100%

Seite 71 - Preparation_

Contents OverviewGetting Started ...Chapter 1Introduction to The CEL-281 ...Chapter 2How it Works ...Chapter 3Preparation for Use ...

Seite 72 - Preparation

A Projected Sound Exposure can be calculated when a sample ex-posure has been obtained over a period that is different from the stand-ard eight hour w

Seite 73 - 4.3 Loading Batteries

In the field of noise measurement, it is important to know the amount oftime spent at a given noise level (known as "on time"), or the amoun

Seite 74 - 4.4 Checking Batteries

10%20%30%40%50%60%70%80%90%100%10%5%Probability%Cumulative%L90L50L10dBLLMAXMINdB91000610065%70 80 90 110 120 130 140Figure 3 Comparison of probability

Seite 75 - (Acoustic Calibration)

1 min. L :Lwith Q = 3 taken over a 1 minutesampling period,10 s L:Lwith Q = 3 taken over a 10 secondsampling period,1sL:Lwith Q = 3 taken over a 1 sec

Seite 76

1 min. slow max: Maximum SPL taken over a 1 minutesampling period with slow response,10 s fast max: Maximum SPL taken over a 10 secondsampling period

Seite 77

3 HOW IT WORKS3.1 Principle of OperationThe CEL-281 consists of a Type 2 sound level meter (IEC 651 and ANSIS1.4 1983) with very fast time weighting,

Seite 78

TimerCALDelayCalibrationFully OperationalCheckBatterySetTime WtgSetFreq Wtg20 secondLevel DoseDistributionProfileTime Info.Press RundBPa h%DoseP%DoseS

Seite 79

calibration delay period, during which the operator can elect to perform acalibration. If no calibration is initiated within this period, the instrume

Seite 80 - 4.6 Setting Clock Functions

able. Equivalent noise levels (L ) are also available, calculated with sev-eral different averaging periods.3.1.5 ProfileThis area calculates and stor

Seite 81

Signals from the microphone are amplified and passed through an A- orC-weighting circuit, as selected by the operator. Both weightings complywith the

Seite 82

Contents ... Page

Seite 83

it into logarithmic form and transmits it into the digital section of the in-strument where all subsequent manipulations are performed.3.2.2 Digital S

Seite 84

a store identity number between 1 and 20,odd numbers show start times, and evennumbers show stop times (Section 3.3.9).All times use the twenty-four h

Seite 85

RUN Either: the instrument is waiting to pass thenext full minute of time before starting to logdata, as described in Section 3.3.4.Or: the instrument

Seite 86

This line may also include an additionalidentifying message: for example or.% Display is showing proportion of elapsed timein under or over range cond

Seite 87

The microprocessor also controls the storage and transmission of calcu-lated and measured data. The 32 kbyte EPROM contains computing andcontrol progr

Seite 88

All two-function keys in the keypad have a toggle operation, i.e. press forone of the functions indicated then press again for the other function,whil

Seite 89

In some operational areas, white keys alsohave an automatic repeat function when helddown.When pairs of keys have to be pressed to implement a command

Seite 90

Number of hourswith display OFFwhile not logging________________6+Number ofhours loggingdata or withdisplay ON≤ 163.3.2 Dose K

Seite 91

Similarly, pressing the key displays the current thresholdlevel, while pressing the key again selects the next from a se-ries of up to five threshold

Seite 92

CautionWhen the instrument is restarted with thekey after data logging has been stopped bypressing the + keys simultaneously,the memory contents are c

Seite 93

Contents (continued) ... Page

Seite 94 - 4.9 Wearing the Dosimeter

against a known acoustic level applied by an external calibratorsuch as the CEL-282. The actual adjustment is in increments of0.05 dB, but the display

Seite 95

When the instrument is stopped, (i.e. the + com-bination has been pressed so that no logging is taking place),and in consequence is still fully operat

Seite 96 - 5.1 Sound Level Meter

While average noise levels are being displayed, pressing thekey allows the results to be obtained from four differentexchange rate values (Q). The val

Seite 97

3.3.6 Frequency Weighting Key(grey two function-key: changes frequency weighting used)While in calibration-delay or calibration modes, when theinstrum

Seite 98

When the instrument is fully operational and the instrumenthas been switched off since the distribution area was last in-spected, pressed alone, this

Seite 99 - 5.2 Noise Dose

mulative percent, while a second press selects dB. Similarly,pressing once selects the lower (PROB%) command anddisplays probability percent, while a

Seite 100 - Obtaining Required Results

grams showing cumulative percent and probability percent areillustrated in Figure 9.3.3.8 Profile Key(beige three-function key: yields profile informa

Seite 101 - Obtaining Required Results_

10 s slow max: Maximum SPL taken over a 10 secondsampling period with slow response,10 s imp. max: Maximum SPL taken over a 10 secondsampling period w

Seite 102 - 5.3 Time Information

1 s slow max: Maximum SPL taken over a 1 secondsampling period with slow response, withresults stored for only the most recent hour,1 s imp. max: Maxi

Seite 103

CautionWhen delayed start and stop times and dateshave been set, ANY changes to the real timeclock disable their operation, and will requirethat all d

Seite 104

Contents (continued) ... Page

Seite 105 - 5.4 Distribution

In any other display mode, this key combination selects theupper (SET) command on the key, cancels the other mode sothat the time and date can be set.

Seite 106

Start times are identified by an odd numbered store location,while stop times have an even number.Start and stop times are displayed in nn:hh:mm forma

Seite 107 - 5.5 Profiles

Then moving the cursor off the display (either to left or right)selects the start or stop date which can also be edited. Tomove from one store locatio

Seite 108

+ (both keys pressed simultaneously)While in calibration delay or calibration modes or when thedisplay has been switched off, this key combination wil

Seite 109

Q = 3, 4, 5, and 6. The key may also be used when bothand keys have been used to change criterion or thresholdlevels in the current calculations.Howev

Seite 110 - 5.6 Example Operations

(yellow key: enables results to be calculated with differentthreshold levels)This key is active only while %Dose, Projected %Dose or Lvalues are being

Seite 111

Under range count,Elapsed time in seconds.Similarly, the following noise information is stored while the instrumentis switched OFF.Histograms (based o

Seite 112

Connection with the CEL-6012 Keypad (or with a PC) is accomplished viathe 15-line connector of the built-in CEL low-power interface bus.1 Keypad2 Seri

Seite 113

All communication is at CMOS logic levels, nominally 0 V and 5.0 V.The connector is located behind a sliding panel in the bottom of the in-strument. F

Seite 114

When the battery voltage falls to 4.2 V, the LOW BAT message ap-pears on the display, and the batteries should be replaced (see Sec-tion 4.3).How It W

Seite 116

4 PREPARATION FOR USEAs the CEL-281 Dosimeter measures and stores all parameters simulta-neously, it can be operated using stored or default settings

Seite 117

+ means while holding down, also press to obtainthe required command. The first key pressed in any combination will bethe first read and acted upon by

Seite 118 - 6 PRINTING HARD COPY REPORTS

Indication of calibration delay mode, displayblinks for 20 seconds.Instrument now operational, shows currenttime weighting and frequency weighting.5.

Seite 119 - DIP Switches

6. When necessary, carefully remove the keypad from the CEL-281by sliding it downwards, until it is clear of the grooves in thesides of the dosimeter

Seite 120 - Caution !

Correct connection and polarity of the batteries is achievedONLY when the holder is inserted with this orientation.6. Replace the battery compartment

Seite 121 - Printing Hard Copy Reports

6. If the instrument is fully operational but the display has beenswitched OFF, press the key to switch the display ON,then go to step 8.7. When the i

Seite 122

Normally, calibration will be undertaken using either a CEL-282 AcousticCalibrator (IEC 942 Class 2L) or a CEL-284/2 Acoustic Calibrator (Class1L), bo

Seite 123

Again, make sure that microphone is correctly seated againstthe shoulder in the base of the cavity, noting that to aid re-moval, the coupler is design

Seite 124 - 7.1 Schedule of Parts

8. When the instrument is logging data (with RUN legend dis-played continuously), press the + keys simultane-ously to stop logging.The instrument reve

Seite 125 - 7.2 Replacing The Microphone

However, the calibration of the instrument may have changedsince it was last checked due to variations in the ambient condi-tions. For altitude and en

Seite 126 - Rear Cover

NOYESAreCurrent Time & DateSetSet Time WeightingAre ThereBatteries In TheCEL-281Load BatteriesIsBattery VoltageOK(Section 6.3)Replace Batteries(Se

Seite 127 - Technical Information_

14. If it is required to change the time weighting in use, follow theinstructions in Section 4.7.1.15. If it is not to be used immediately, switch the

Seite 128 - Technical Information

2. If the instrument is switched OFF, switch the power ON asdescribed in Section 4.1.The ON legend will appear on the display.3. If the CEL-6012 Keypa

Seite 129

8. When the instrument is logging data (with RUN legend dis-played continuously), press the + keys simultane-ously to stop logging.The instrument reve

Seite 130 - 7.3 Specification

14. Press the and keys to move the blinking cursor alongthe line of characters to the left or right to select another digitfor editing.Moving beyond t

Seite 131

All entries must be made in order: i.e. entries in higher numbered storelocations must be for later times and dates. Once the entry in the laststore l

Seite 132

6. Press the and keys to change the indicated digit oneunit at a time.7. Press the and keys to move the blinking cursor alongthe line of characters to

Seite 133

With the following formula, an estimate can be made whetherthere will be sufficient battery life at room temperature to per-form any particular loggin

Seite 134

4.7 Making Preliminary SettingsWhen the dosimeter is switched ON and operated, it will use either thedefault settings set at the factory or settings r

Seite 135

4. If the instrument is already fully operational but the display hasbeen switched OFF, press the key to switch the displayON, then go to step 8.5. If

Seite 136 - WARRANTY ARRANGEMENTS

(Impulse) symbols (see Section 3.2.2), depending on whichtime weighting is currently set.9. Press the or keys to change the time weighting to therequi

Seite 137 - Service & Warranty_

1 GETTING STARTEDIt is not necessary to re-program theCEL-281 each time it is used !In fact, if the settings for frequency weighting and time weightin

Seite 138 - Return Information

5. If the instrument is fully operational while the RUN legend isdisplayed continuously to indicate that data is being logged, goto step 6, but note t

Seite 139

4.8 Starting And Stopping Data LoggingIn order to generate noise distribution histograms, calculate noise dosevalues, and produce noise profile time h

Seite 140 - Reference PageReference Page

4. If the instrument is in calibration delay mode or calibrationmode, if necessary perform a field accuracy check (acoustic cali-bration) according to

Seite 141

The RUN legend is displayed, and the instrument will start tostore samples of data at the beginning of the next whole m-inute. The instrument continue

Seite 142

The RUN legend is extinguished and the instrument reverts tofully operational mode with no data being logged.All preliminary set up operations and com

Seite 143

Preparation_

Seite 144

5 OBTAINING REQUIRED RESULTSProvided the CEL-281 Computing Noise Dosimeter has been preparedfor use as detailed in Chapter 4, it will store ALL parame

Seite 145

If the instrument has just been calibrated or the 20 secondcalibration delay period has elapsed, it will enter the soundlevel meter measurement area a

Seite 146

5. Provided the instrument is not logging data, if necessary, pressthe key to change the frequency weighting in use (see Sec-tion 4.7.2).6. While soun

Seite 147

8. While average noise levels are being displayed, press thekey to show the currently used threshold level on the main dis-play.9. If necessary, press