Casella-cel CEL-278 Bedienungsanleitung

Stöbern Sie online oder laden Sie Bedienungsanleitung nach Audio-Zubehör Casella-cel CEL-278 herunter. Casella CEL CEL-278 User Manual Benutzerhandbuch

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 34
  • Inhaltsverzeichnis
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 0
CEL-278/2 OCTAVE FILTER SET
Operator's Handbook
PP60024
Issue: 2
August 1990
CEL Instruments Ltd. CEL-278/2 Handbook Page 1
Seitenansicht 0
1 2 3 4 5 6 ... 33 34

Inhaltsverzeichnis

Seite 1 - Operator's Handbook

CEL-278/2 OCTAVE FILTER SET Operator's Handbook PP60024 Issue: 2 August 1990 CEL Instruments Ltd. CEL-278/2 Handbook Page 1

Seite 2

On "Hi" range the input signal passes directly to the anti-aliasing filter and thence to the selected octave filter. On "Lo" range

Seite 3 - Contents Page

Figure 2 Typical octave filter response curve pressed, or band step control signals are received via the bottom connector from an external device su

Seite 4

battery voltage falls to an unusable level, and whether an overload condition is occurring. Figure 3 shows the full annotation provided by the display

Seite 5 - 1. INTRODUCTION

Figure 3 Control panel and display showing all indicators (Normally, all indicators will not be displayed at the same time) placed, or the external

Seite 6

Normally when the Filter Set is used with CEL sound level meters, (except for the CEL-493/2 series), the CEL-278/2 RANGE switch is set to "Hi&quo

Seite 7 - 2. SCHEDULE OF PARTS

When the button is pressed while standing in the highest frequency filter band, the instrument returns to the 'All Pass" band. Conversely

Seite 8

1. Conditioned AC signals up to 8.0V RMS, 11.3Vpk with a minimum load impedance of 10kΩ output via line 6 of the top connector, and line 4 of the bot

Seite 9 - 3. DESCRIPTION

3.6 Batteries & Auxiliary Power Supplies The CEL-278/2 is designed to operate with the set of four manganese alkaline batteries (IEC Type LR6 (AA

Seite 10

Incorrect installation, handling, operation, and disposal of batteries can be dangerous. Follow the battery manufacturer's recommendations, and i

Seite 11 - 3.2 Display

4. PREPARATION FOR USE The CEL-278/2 is fully tested and calibrated in accordance with the requirements of its specification before it is dispatched

Seite 12

Page 2 CEL-278/2 Handbook CEL Instruments Ltd

Seite 13 - 3.3 Control Identities

When a set of batteries has been loaded, check their power output by sliding the RANGE switch to "On" to ensure that LOW BAT is not indicate

Seite 14

5. OPERATION The Filter Set can be used for frequency analysis only when connected to a suitable measuring instrument, and both must have power avail

Seite 15 - 3.5 Output Facilities

Figure 7 Attaching the filter set close coupled to a CEL sound level meter Ensure that the locking screw is correctly aligned with the bush in th

Seite 16

Figure 8 Interconnection with other CEL measuring and recording instruments CEL Instruments Ltd. CEL-278/2 Handbook Page 23

Seite 17

Warning Remove batteries from both Filter Set and Sound Level Meter if they are to be powered by an external supply. Alternatively, connect the ins

Seite 18

5.2 Manual and Remotely Scanned Octave Band Frequency Analysis When the CEL-278/2 is connected to a sound level meter, any noise (or vibration) param

Seite 19 - 4. PREPARATION FOR USE

The Instruments are now set so that the instructions given in the level recorder handbook can be followed to set up and record spectrums, with freque

Seite 20

6. SPECIFICATION Filter bands: 11 Octave filter bands according to IEC 225 and ANSI S1.11-1966 (R1975) Class II, C-weighted network according to Typ

Seite 21 - 5. OPERATION

Typical noise floor: (measured on Lo range with a CEL-493/2 Sound Level Meter) Filter Set All pass: (WTNG and band position) Linear A-weighted 2

Seite 22

Max output voltage: 8.0V RMS signal (26Vpk-pk under overload conditions), Output impedance: <1kΩ, Recommended input impedance of output device 10

Seite 23

Contents Page 1. Introduction 5 2. Schedule of parts 7 3. Description 9 3.1 Technical description 9 3.2 Display 11 3.3 Control identities

Seite 24

External power, 3.5 to 8V DC via pins 1 and 2 of auxiliary socket, at 250mA maximum. Typically 100mA at 6V. Batteries: 4 x IEC LR6 Alkaline Mangan

Seite 25 - Analysis

7. MANUFACTURERS SERVICING & WARRANTY ARRANGEMENTS In order to ensure its rigid conformity with the requirements of the specification, the instru

Seite 26

Page 32 CEL-278/2 Handbook CEL Instruments Ltd

Seite 27 - 6. SPECIFICATION

8. CALIBRATION CERTIFICATE The CEL-278/2 Octave Band Filter Set described below has been inspected andua!ih/u1edinaocordunoevvbh1herequi,onlentony1ho

Seite 28

Page 34 CEL-278/2 Handbook CEL Instruments Ltd CEL Instruments are designed, manufactured, and serviced by: Lucas CEL Instruments Lt

Seite 29

Page 4 CEL-278/2 Handbook CEL Instruments Ltd

Seite 30

1. INTRODUCTION The CEL-278/2 Filter Set was designed to meet the requirement for a portable filter set that allows octave frequency analysis to be pe

Seite 31 - ARRANGEMENTS

The CEL-278/2 may be used with other types of CEL instrumentation, including the following. CEL-160 Graphic Level Recorder (not "C" version

Seite 32

2. SCHEDULE OF PARTS A complete "CEL-278/2 Octave Filter Set" contains the following items. When the instrument is delivered, check that al

Seite 33 - 8. CALIBRATION CERTIFICATE

Page 8 CEL-278/2 Handbook CEL Instruments Ltd

Seite 34

3. DESCRIPTION 3.1 Technical Description The CEL-278/2 is designed to be mounted close coupled to the bottom of a CEL sound level meter. It receives

Kommentare zu diesen Handbüchern