10.2 Additional Instructions for Use With
B & K 1621 Tunable Band Pass Filter
10.2.1 General
10.2.2 Connections
10.2.3 Operation
CEL-160 Control Settings
a) POWER On
b) CONT- Off
c) EXP TRAC E - Off
d) GAIN - Cal
e) RANGE- 120
f) DETECTOR- RMS
g) TIMECONSTANT- Fast
h) RECORDING - Log
i) WEIGHTING- Filter
J) DYNAMIC RANGE 50 dB
B & K 1621 Control Settings
a) FILTER- 3%
b) FREQUENCY RANGE- 2 HZ - 20 kHz for
7F,
20 Hz - 20 kHz
for 6F
c) SWEEP CONTROL - Auto
d) MAX VOLTAGE - 0.2 V RMS
Sweep Initiation
a) Enter Filter code 6F or 7F.
b) Enter commencing paper speed 5B for 6F, 7B
for 7F (see para 5.3).
c) Depress Key E. CEL-160 advances to start
position.
d) Depress SWEEP CONTROL to start and
release. The Auto LED will be lit until the
decade sweep is completed.
e) Advance FREQUENCY RANGE to the next
decade and repeat step (d) when the AUTO
LED is extinguished. Repeat until the sweep is
complete.
10.2.4 Calibration Considerations With
Alternative Signal Inputs
CEL-186/2980/3686 Input
CEL-3025/2980/3686 Logarithmic
Recording
CEL-3025/2980/3686 Linear Recording