12. EQUIPMENT SPECIFICATION
SPECIFICATION CHARACTERISTICS
STANDARDS
IEC
61672-1
2002-5 (Electro-Acoustics
–
Sound
Level
Meters)
Gro up
‘X’
instruments,
Performance
Class
2.
IEC
60651: 1979 Type 2 .ANSI S1.4
type
2A Specification for
Sound Level Meters.
CEL-244/246 Models:
IEC 60804: 2000 Type 2, ANSI
S1.43: 1997 (R2007) Type 2
RANGE:
Display
range:
30 -130dB(A)
RMS,
available
in
2
ranges,
30 -100dB
and
60 -130dB.
Linear
operating
range
10dB
above
noise
floor.
RMS
FREQUENCY
WEIGHTINGS
A
and
C
filter
weightings,
satisfying
IEC
61672-1:
2002
Clas s
2,
ANSI
S1.4
Type
2A.
NOISE
F
LOOR
Total
noise
floor
typically
<33dB(A).
FREQUENCY
RESPONSE
Overall
frequency
response
as
per
IEC
61672-1:
2002
Class
2,
ANSI
S1.4
Type
2A.
TIME
WEIGHTINGS
Fast,
Slow
and
Impulsive
according
to
IEC
61672-1:
2002,
ANSI
S1.4
Type
2A.
REFERENCE
CONDITIONS
20
o
C
air
temperature,
65%
relative
humidity,
101.325kPa
atmospheric
pressure. Nominal
reference
level
=
114.0dB
at
1k Hz.
Free
field
perpendicular
incidence.
OPERATING
ENVIRONMENTAL
CONDITIONS
Humidity
5
to
90%
RH
in
the
absence
of
condensation.
Temperature Range
0 – 40
o
C.
Pressure
65
to
108kPa.
Effects
of
Humidity
Less
than
±0.5dB
over
the
range
30
to
90%
relative
humidity
(non-condensing),
relative
to
th e
value
at
reference
conditions.
STORAGE
ENVIRONMENTAL
CONDITIONS
Humidity
0
to
90%
RH
in
the
absence
of
condensation.
Temperature
Range
-20
to
+60
o
C.
Pressure
65
to
108kPa.
MICROPHONE
Internal
electret
capsule
10mV/Pa
nom
+/-3dB
within
½”
preamplifier.
CALIBRATION
Auto
calibration
by
app lication
of
1kHz
@
114
or
94dB
calibrator
(CEL-110/2
or
CEL-110/1).
±1dB
calibration
range.
POWER
SUPPLY
External
DC
5VDC
(via
US
B
5
pin
mini
B).
Batteries
3x
AA,
Rechargeable
NiMH
or
Alkaline
cells.
17