4. Replay the recorded calibration signal.
5. Adjust the Replay Level control on the recorder so that the sound
level meter display indicates the calibration level for the
microphone and response used.
6. Stop the recorder.
The system is now calibrated to replay at accurately determined
levels with the calibration signal near the top of the input range,
which gives the widest possible replay dynamic range. However,
the instrument measurement range may need to be adjusted so
that the display scale gives a correct indication of the recorded
values.
7. Press to enter quick edit mode.
8. Use and to select the measurement range that was noted
for the measurement.
9. Press again to confirm the calibration and any scale change.
The instrument replay scale now matches the scale used for
the recording so that the levels indicated by the display will be
correct.
10. Press to display the measurement screen, then replay the
recorded material and perform any measurements and analysis
required.
MENU
dB
dB
Ø9Jan98 12.22
Scale 5Ø-12Ø
L
ZF
k-20
Operation