8. DIGITAL INTERFACE
Digital interfacing is achieved via the 18-way connector of the auxiliary
socket (illustrated in Figure 5. The following table shows the line use.
Table 1 Line use in the 18-way Lemo Connector
Pin N
o
Name Centronics Use RS 232 Use Other Use
1 INIT
Event Active Event Active Event Active
2 POUT2
Data 3
Not Connected Unused
3 AC OUT Not Connected Not Connected
AC Out
4 0 V 0 V Ground 0 V Ground 0 V Ground
5 +5 VL Not Connected
+5 V +5 V
6 DC IN Not Connected Not Connected
Ext Power
6 - 15 V DC
7
STROBE
Data Strobe
Not Connected Unused
8 POUT1
Data 2
DCD
Unused
9 LOG DC Not Connected Not Connected
Log DC
10 P BUSY
Busy
CTS
Unused
11 POUT6
Data 7
RTS
Unused
12 POUT7
Data 8
DTR
Unused
13 SPARE2 Not Connected Not Connected Not Connected
14 POUT3
Data 4
RXD
Unused
15 POUT0
Data 1
DSR
Unused
16 SPARE1 Not Connected Not Connected Not Connected
17 POUT5
Data 6
Not Connected Unused
18 POUT4
Data 5
TXD
Unused
Notes: Signals shown with are outputs from the 268/368,
Signals shown with are inputs to the 268/368,
Signals shown with are birectional controlled by the 268/368 software.
The spare Lemo pins 13 & 16 are currently not connected in the 268/368 but
may be connected in future versions.
RS 232 pins 10 & 11 are linked in the Lemo plug (
RTS & CTS) and this link is
tested when determining whether the printout goes to a Centronics or RS 232
interface.
Pins shown in "Other Use" column as "Unused" are buffered or unbuffered CPU
port lines. Outputs will be either 0 V or +5 V CMOS Logic.
CEL-268/368 Handbook