APEX SERIES
PERSONAL AIR SAMPLING PUMPS
& PUMP MANAGER SOFTWARE
(I.S. and Non I.S. Versions)
User Handbook
HB 3294-06
COPYRIGHT
The copyright in this document which contains proprietary information is vested in CASELLA CEL LIMITED. The
contents of this document must not be used for purposes other than that for which it has been supplied or
reproduced or disclosed wholly or in part without the prior written permission of CASELLA CEL
CASELLA CELCASELLA CEL
CASELLA CEL
CASELLA CEL
Regent House
Wolseley Road
Kempston
Bedford
MK42 7JY
Tel: 00 44 1234 844100
Email: info@casellameasurement.com
www.casellameasurement.com
Casella USA
17 Old Nashua Rd
Amherst
NH03031
USA
Tel: 00 1 800 366 2966
Email: info@casellausa.com
www.casellausa.com
Seitenansicht 0
1 2 3 4 5 6 ... 39 40

Inhaltsverzeichnis

Seite 1 - HB 3294-06

1 APEX SERIES PERSONAL AIR SAMPLING PUMPS & PUMP MANAGER SOFTWARE (I.S. and Non I.S. Versions) User Handbook HB 3294-06 COPYRIGHT The co

Seite 2

10 Apex Pro I.S. (Intrinsically Safe version of the Apex Pro). Please consult Casella CEL for information about upgrading a standard model to an Ape

Seite 3

11 1.2 Display Symbols Halt - shows when pump is in Hold Mode , or blinks when halt option available. Run - shows when pump motor is running.

Seite 4

12 DO NOT open the battery compartment of I.S. versions in a hazardous area. DO NOT charge I.S. battery packs in a hazardous area. The battery pac

Seite 5

13 The following additional instructions must be read when using Apex Lite Pump versions. WARNINGS ! Apex Lite Versions are not intrinsically sa

Seite 6

14 TEMP Current internal air temperature. 2.2.2 Manual Reset (Clear Current Sample) To clear any accumulated sample duration and volume, the ins

Seite 7

15 02041FlowFlowOutletApex PumpCycloneSampling HeadDry Cal FlowmeterConnecting tube (5 mm nominal internal dia.)Dry Cal®Primar y Flow Mete rTE MP

Seite 8

16 ESP Spanish, DAN Danish. TEMP Sets the temperature display units asCentigrade or Fahrenheit. VOL Sets the volume display mode to Auto or m

Seite 9

17 the buzzer will sound. When the problem lasts for three or more seconds, the pump motor will switch itself OFF and remain OFF with the instrument

Seite 10 - TEMP 19.48

18 must be sampled for a long period. The duration of the run can be set from the instrument keypad. On Apex Pro, the duration can also be set via

Seite 11

19 For example, when 10 litres must be sampled over an 8-hour work shift, the pump flow can be set to run at a flow rate of: = 0.021 litres / min =

Seite 12 - Ni-MH Battery Pack

2 WARNINGS ! All Versions: Apex and Apex Pro air sampling pumps are designed to be robust, however they should not be dropped or subjected to mechan

Seite 13 - 2.2.1 Hold Mode

20 8. Press any key to return to Hold Mode. The whole duration is stored as a single “sample” (run) with each ON / OFF sequence of the pump include

Seite 14

21 A display message asks for the upper calibration point to be set. 6. Use and to set the flow required at the calibration point. The ins

Seite 15 - 2.4 Configuration Mode

22 Pump Manager software automatically detects the language of the operating system during the installation. Languages supported include: English

Seite 16

23 the bottom of the PC display. This indicates that the device is waiting to communicate with an Apex Pump. Best communication with the pump is a

Seite 17 - Primary Fl ow Meter

24 3. Select the File menu and see the following options. New This option allows pump and person details to be added to the Directory field in

Seite 18

25 of an existing pump) to display its details. (A New Pump dialog is displayed automatically whenever the soft-ware detects an unknown pump within

Seite 19 - = 10 minutes

26 6.5 Downloading Data DO NOT use the IR communication link for I.S. versions while in a hazardous area. Data is downloaded from pump to PC autom

Seite 20

27 If a pump is detected that has not been configured by the software, i.e. is not included in the Pump Manager Directory field, a “New Pump” mess

Seite 21

28 1. Display data by highlighting the name of the relevant pump or person in the Directory. A directory containing all samples from the selected p

Seite 22

29 Analysis Page adds analysis information to the sample folder. Sample number A sample identity can be added, Sample data Sample analysi

Seite 23

3 location where the external conditions are conducive to the build-up of electrostatic charge on such surfaces. Additionally, the equipment shall o

Seite 24

30 Date The date of the sample or of the report can be entered, Audited by The data / report checker can be identified, Date The date o

Seite 25

31 As shown on the fold out sheet at the back of this book, each program has its own individual Properties dialog for information specific to the pr

Seite 26 - 6.5 Downloading Data

32 13 Specify Start and End dates and times on the Program Times dialog. 14. Enable the selected Time (Ö) so it will be used by the program. 15.

Seite 27 - Printing a Report

33 with safety time out. Standby trickle charge mode to keep pump ready for use. Typical charge time ( N/A for Apex Lite). 3 hours for full

Seite 28

34 7.5 I.S. Certification

Seite 29

35 7.6 EC Declarations of Conformity

Seite 31

37 8. SERVICING All servicing on I.S. versions MUST be undertaken by returning the unit to the Casella CEL Service Department. Casella CEL’s in h

Seite 32 - 7. TECHNICAL INFORMATION

38 182094A Dry-Cell Battery Pack for use only with non-I.S. versions and Apex Lite. P104104 Standard Starter Kit, includes standard carry case wit

Seite 34

4 besteht.Zusaetzlich ist das Geraet mit einem feuchten Tuch zu reinigen AVERTISSEMENTS ! Toutes les Versions: Les pompes de prélèvement d’air Apex

Seite 36

5 certains produits chimiques. Dans certaines circonstances extrêmes, les pièces non métalliques incorporées dans la fermeture de cet appareil peuve

Seite 37

6 entonces es responsabilidad del usuario tomar las precauciones pertinentes para prevenir efectos adversos que pudieran afectarlo, de forma que se

Seite 38

7 utilizzati per la sua costruzione: Policarbonato ad impatto elevato PC- ABS/ trasparente Nel caso probabile in cui l’apparecchiatura possa venire

Seite 39

8 CONTENTS 1. INTRODUCTION 9 1.1 Control Keys 10 1.2 Display Symbols 11 1.3 Indicators & Display Messages 11 2. PREPARATION FOR USE

Seite 40

9 Improved Features ¤ Small size due to space-saving pump design ¤ Real-time display of flow rate on large LCD ¤ Direct flow control via keypad ¤