5. Perform Acoustic Calibration
Operations are performed by a sequence of key strokes.
Perform a field accuracy check
each time the instrument is switched ON
using a CEL-282 (Class 2) or CEL-284/2
(Class 1) Calibrator as follows.
Fit the microphone into the CEL-
4725 Coupler supplied with the calibrator,
making sure it is pushed firmly into contact
with the shoulder in the coupler cavity.
Then fit the coupler complete with micro-
phone into a CEL-282 (or CEL-284/2) Cali-
brator, again ensuring that it makes contact
with the shoulder in the calibrator cavity.
(To aid removal, the coupler flange does
not fit close against the calibrator housing).
Switch the calibrator ON, then fol-
low the instruction after the Start Up sequence of displays has finished.
Calibrator
980057
1/4"
Microphone
O'Ring
Shoulder
Microphone
Coupler